OPTI-SEAL® UHMW-PE Piston Seal for Dionex ICS-3000 and ICS-5000 analytical pumps is a precision-machined replacement piston seal for Thermo/Dionex OEM part number 064946, also commonly searched as 64946.
This seal is designed for ICS-3000 and ICS-5000 analytical versions only. It is not intended for capillary pump versions.
Optimize Piston Seals are among the longest lasting seals on the market. Machined from proprietary polymer blends, Optimize Seals provide excellent high pressure sealing performance. For mostly aqueous or buffered mobile phases, OPTI-SEAL® UHMW-polyethylene is the recommended choice. Specially designed to be more resistant to abrasion and wear than standard polyethylene seals, they can last up to seven times longer under mostly aqueous conditions.
Despite UHMW-PE's durability, it has a coefficient of friction approaching that of PTFE. The homogenous polymer nature leads to less shedding, and when shedding does occur, the larger, more fibrous particles are more easily captured by a filter frit. For a specific recommendation on seal material for your mobile phase, consult our Solvent/Seal Compatibility Table.
| OEM Part Number | 064946 |
|---|---|
| Alternate OEM Search Format | 64946 |
| Product Type | Main Piston Seal |
| Compatible Systems | Dionex ICS-3000, ICS-5000 |
| Pump Version | Analytical Version Only |
| Material | UHMW-PE / UHMW-polyethylene |
This piston seal is for ICS-3000 and ICS-5000 analytical pump versions only. It is not intended for capillary pump versions. Please confirm your system configuration before ordering.
| Product Type | Piston Seal |
|---|---|
| Seal Function | Main Piston Seal |
| Material | UHMW-PE / UHMW-polyethylene |
| Recommended Conditions | Aqueous or mostly aqueous mobile phases |
| OEM Cross Reference | 064946 / 64946 |
| Compatible Systems | Dionex ICS-3000, ICS-5000 analytical pumps |
A worn or damaged piston can prevent a new seal from seating correctly and may cause premature leakage. If the piston is scratched, hazy, flattened, or visibly worn, replacing the seal alone may not solve the problem.
Yes. This OPTI-SEAL® UHMW-PE piston seal is designed as a replacement for Thermo/Dionex OEM main piston seal 064946, also commonly searched as 64946, for compatible analytical pump systems.
No. This seal is for ICS-3000 and ICS-5000 analytical versions only. It is not intended for capillary pump versions.
UHMW-PE is recommended for mostly aqueous or buffered mobile phases. For mostly organic mobile phases or applications requiring broader solvent compatibility, another seal material may be more appropriate. Please consult the Solvent/Seal Compatibility Table for guidance.
Immediate leakage can be caused by installation damage, a dry seal, improper seating, or a worn or scratched piston. Inspect the piston carefully and confirm that the seal was wetted or pre-soaked before installation.
If the piston shows signs of wear, scratches, haziness, flat spots, or salt buildup, it should be replaced. A damaged piston can shorten seal life and cause leaks even when a new seal is installed correctly.
Depending on your system condition and service procedure, you may also need the piston seal wash seal, seal insertion tool, O-ring, piston, or other preventive maintenance components.