HPLC/UHPLC Trap Columns

Trapping - What is it? What are trap cartridges and how are they used?

Trapping: The selective retention and subsequent elution of analytes within a sample. Trapping is a chromatographic technique, but it does not exploit typical phase-boundary effects such as partitioning. Many trapping applications could be described as "digital" or on/off chromatography, because solvent conditions are selected to ensure that there are only two retention states - one where solvent strength is weak enough to allow the analyte to bind to the stationary phase without eluting, and one where solvent strength is sufficient to cause immediate and complete sample elution. The term trap cartridge refers to a packed bed with sufficient capacity to completely retain a given amount of the target analyte in a sample. Trapping is often used in conjunction with mass spectrometry - the trap cartridge serves to clean up (and/or pre-concentrate) the analyte, while the MS handles detection and identification. The MS can also compensate for low separation resolution and column efficiency that would be unacceptable with other detection modes. In return for the trade-off in resolution, trap cartridges offer greatly reduced analysis time (a benefit common to all short/fast bed formats).

 

Example 1: Sample Cleanup ("Trap and Dump"), Desalting

The analyst has a simple mixture - a compound of interest dissolved in a medium not suitable for injection onto an MS (highly aqueous, salt-containing, etc.). For instance, protein samples often arrive for analysis in buffered solutions. This might be for stability reasons, or the salts might be present due to a preparation step such as a digest. This salt has to be removed before mass spectrometry analysis, as it will foul the MS interface. A trap cartridge can be placed within the loop of an injection valve to capture the analyte conveniently. Once the buffer salts and impurities have been washed away, mobile phase strength can be increased, and the compound can be eluted onto an analytical column or directly onto the MS.

 

Example 2: LC Separations on the MS Timescale

A big problem with LC-MS is that the LC part takes a lot longer than the MS part. With MS, it is not always necessary to achieve high resolution, but a trap cartridge can be used to provide a small amount of chromatographic separation. Using switching valves, two trap cartridges could even be used in a tandem/parallel set-up, reducing the LC cycle time by 50%. These two factors can be combined to produce an automated, fast LC-MS system that runs at the pace of the MS, not the LC.

 

Example 3: Separation of A Mixture of Proteins

Traps are often used in protein separations. You can actually use two or more cartridges in a row to assist with the separation of mixtures of charged and neutral proteins. For instance, an ion exchange cartridge could be used in-line, followed by a C18 cartridge. Additional switching and injection valves would be used to offer increased solvent routing flexibility, as follows:

Gradient pump → Injection Valve → Ion Exchange Trap → Switching Valve → C18 Trap → Mass Spectrometer

The protein or peptide mixture would first be injected inline ahead of the ion exchanger. Using a weak mobile-phase solvent would cause charged proteins to be retained by the ion exchanger, while neutral proteins would pass through to the C18 cartridge, where they would be retained. A switching valve between the exchanger and the C18 bed allows switching the solvent delivered to the C18 trap only. Neutral proteins can be eluted from this trap onto the MS by increasing the solvent strength. Then, charged proteins could be selectively eluted from the exchanger onto the C18 column via a salt gradient (the valve downstream of the C18 trap would switch so that salts were directed to waste rather than to the MS). A second subsequent elution off the C18 trap (in conjunction with a solvent switch) sends these charged proteins onto the MS, free of salts and contaminants.

 

Optimize Products for Trapping

Optimize offers a wide selection of novel packed-bed products, all of which can be used effectively in a sample trapping or on-line purification application. The following products are well-suited for use as trap cartridges. Still, we encourage you to call us to discuss your particular requirements - we have numerous bed formats in a variety of dimensions, and many options for standard or custom-packed stationary phases that can be tailored to your application. Our trap cartridge offering is centered on four main products with many variations: a low-volume capillary product suitable for direct installation into injection valve ports and columns; a variable-volume quick-connect product that can be connected in-line anywhere on your system; a low-volume ultra-high-pressure stem trap; and a variable-volume ultra-high-pressure hand-tight cartridge trap.

 

OPTI-PAK® Capillary Traps

This product is perfect for direct connection to injection and switching valves and offers a very low-volume bed format, combined with an auto-adjusting ZDV connection to any 10-32 port. Use for very small sample volumes and low flow rates. This product is rated to 6,000 psi. 

OPTI-LYNX™ Trap Cartridges

These cartridges offer quick-connect convenience and a larger bed for greater sample capacity and higher flow rates.

EXP® Stem Traps

The slim architecture allows it to fit easily into crowded instrument compartments or connect directly to tightly spaced injection ports. When tightened by hand, it seals to 8,700+ psi. When tightened with the incorporated wrench flats, it seals up to 20,000+ psi. The low-volume, low-dispersion cartridges are perfect for your UHPLC volume-critical applications.

EXP® Trap Cartridges

The Hand-Tight EXP Trap Column is rated for use at 20,000+ psi and can connect to any injection valve or column with 10-32 threads or can be connected in-line with 1/16” stainless tubing. For your UHPLC applications that require a larger packed bed, these cartridges are the best choice on the market today.

HPLC/UHPLC Guard Columns

When do I need to use a Guard Column?

Guard columns perform two primary functions:

  • First: they capture irreversibly retained compounds - substances that would otherwise permanently bind to and foul the stationary phase of the analytical column. They guard the analytical column from premature failure.
  • Second: they serve a sacrificial role - when mobile phase pH is extreme enough to degrade the column packing, a guard column with similar phase characteristics can "take the hit", and prolong the life of the analytical column. Guard columns also function as pre-column filters, protecting the analytical column from particulate contamination. However, if the mobile phase is at a moderate pH and the samples are free of irreversibly retained compounds and other contaminants, particulate contamination can be handled without a guard column, using only a precolumn filter.

 

Selecting a Guard Column: Factors to Consider

Liquid chromatography has become an increasingly sophisticated separation technique over the last two decades, as advances in instrumentation have led to lower detection limits, greater automation, and more powerful data-handling capabilities. Similarly, improvements in packing materials and column design have led to more efficient, selective, and reproducible separations. Even with these substantial design improvements, the analytical column remains a problematic link in the HPLC chain. With the cost of quality analytical columns continually increasing, and with the price of some specialty analytical columns reaching into the thousands of dollars, effective pre-column protection is a wise precaution in today's cost-conscious HPLC laboratory.

Choosing a Guard: Several factors are important to consider when selecting a guard column. The fundamental role of a guard column is to trap particulates, chemical contaminants, and other foreign material, preventing them from fouling the analytical column. A guard column with the same packing material as the analytical column will provide the best protection. Chemical contaminants and other substances with a propensity for fouling octadecyl stationary phases would be best trapped by an octadecyl guard column; contaminants that cause deterioration of a particular packing material would first attack the stationary phase in the sacrificial guard, protecting the analytical column for as long as the capacity of the guard column will allow.

A second factor to consider is a guard column's potential to contribute to dispersion by introducing extra-column volume. The installation of a guard column requires additional connections and often extra tubing to be incorporated into a section of the chromatograph where excessive dead volume can make or break an assay. A well-designed guard column will minimize the amount of extra tubing required for connection to the analytical column, thus minimizing extra-column dispersion.

When selecting optimal precolumn protection for your HPLC, you should also consider how the guard column may affect separation efficiency. Using a guard column extends the packing bed by a finite distance. A well-designed guard column will impart minimal influence on the efficiency of the analytical separation. A guard column that significantly affects efficiency will cause problems when validation time arrives or when the guard itself must be replaced.

Optimal Protection by Design:  If a prospective guard column is to provide optimal performance and protection for your analytical column, it must address the physical and chemical constraints previously mentioned. Guard Columns from Optimize have been designed to provide reliable, low-dispersion column protection while eliminating the risk of dead volume in the precolumn region. Optimize Guard Columns offer a tool-free connection with an automatically adjusting tube stem for perfect zero-dead-volume connections, regardless of the analytical column manufacturer's tube stop depth. This patented design guarantees that the tube stem will automatically bottom out securely against the tube-stop of any 10-32 female port. Some guard columns incorporate molded one-piece couplers or swaged-on ferrules that establish a fixed tube-stop depth for connection to the analytical column. Guards with fixed tube stems will introduce dead volume or leaks at the connection to the analytical column when switched from one port to another. This situation can occur even when a fixed-stem guard column is switched between different 10-32 ports from the same manufacturer. Because tube-stop depth varies from port to port due to normal manufacturing tolerances, a guard column with a fixed tube stem will often leak or introduce dead volume when switched between different female 10-32 ports.

A guard column should perform its primary function of column protection in a manner that is as chemically transparent as possible to the chromatographic separation. When selecting a guard column, choose high-quality packing materials that are well-matched to the stationary phase of your analytical column. Also, be sure that the guard column you select is designed to minimize extra-column effects. Guards that avoid fixed or pre-swaged tube stems and connect directly to the analytical column, without extra tubing, are the optimal choice.

 

Filtration's Critical Role in Protecting Your HPLC/UHPLC

Filtration: A Systemic Approach to Protecting Your System

Obtaining reliable, consistent results from your HPLC system requires ongoing attention to chemical and mechanical factors that can adversely affect system performance. For your laboratory to remain competitive, instrument downtime must be kept to an absolute minimum. Many common instrument problems are caused by particulate contamination introduced by the mobile phase or generated as a by-product of system operation. Particulates can cause a variety of maintenance headaches, including check valve failure, excessive wear on the piston and seal, and high back pressure due to frit blockage. A consistent, holistic approach to filtration is an effective way to minimize particulates in your flow path. It is also a vital precaution that will keep your chromatograph working for you (and not the other way around).


Mobile Phase / Solvent Filtration: The first and perhaps easiest place to prevent particulate contamination is in the mobile phase. Buffer salts, improperly cleaned glassware, microbial contamination, and pre-filtering procedures are all potential sources of particulates. Solvent manufacturers adequately filter HPLC-grade solvents before bottling; ideally, there should be no need for additional pre-filtering in your laboratory if your mobile phase contains only HPLC-grade solvents. Mobile phases prepared with buffer salts or other solid reagents should, however, be filtered before use. A 0.2- or 0.45-µm membrane filter is recommended for this filtration step. It is also a good idea to discard the first few milliliters of solvent that pass through the filter. Solvent reservoir filters are a useful filtration accessory, providing a first line of protection as the solvent enters the system. OPTI-SOLV® Solvent Reservoir Filters are an excellent choice for this application; OPTI-SOLV secures your inlet tubing at the bottom of the solvent reservoir without tools or fittings and is machined from PTFE and Titanium. This well-swept filter ensures access to all but the last few milliliters of solvent and will not trap bubbles or leach extractable contaminants. Solvent reservoir filters are available with a choice of frit porosities. For analytical HPLC applications, a 2 µm frit is the best choice, while a larger-porosity frit, such as 10 µm, should be selected for flow rates exceeding 10mL/min.



Piston Seals:  Piston seals are a common source of particulate contamination in the flow stream. As the piston moves back and forth through the seal at high pressure, seal fragments can slough off from the seal's mating surface. The size and shape of the particles will vary depending upon the nature and quality of the material from which the seal is made. These variables are important to consider because they directly affect how easily shedding particles can be removed from the flow path. Particles too small to be trapped by a 2 µm filter, or large enough to be trapped within the frit but below the surface, will cause the most trouble. The exclusive polymer blends used in our ITB™ and OPTI-SEAL®  piston seals are specifically formulated to shed particles large enough to be trapped at the surface of a 2 µm frit. Because particulates in the pump head can cause increased friction and abrasion between the piston and seal, excessive seal wear is both a source and a symptom of particulate contamination.


In-Line Filtration:  The buildup of precipitated salts and other particulates within the pump can be minimized by regular flushing with a non-buffered mobile phase. Some pumps include a piston flushing utility that allows the piston to be rinsed on the low-pressure side of the seal while the pump is in use. Daily piston flushing can help prevent salt precipitation on the piston surface and prolong the life of both the piston and its seal. For the most effective protection against seal-shedding particulates, we recommend an in-line filter installed downstream of the pressure transducer and upstream of the injection valve or autosampler. OPTI-SOLV In-Line Filters are a convenient and cost-effective solution for most in-line filtering applications, providing rapid access to replaceable frits without requiring disconnection of attached tubing.



Check Valves with Filters are a bad idea
:  Some instrument manufacturers supply check valve assemblies equipped with a filter element on the outlet side. While a filter frit at this location may seem like a better option than no filter at all, this arrangement is problematic. First, the use of an in-line filter upstream of the pressure transducer complicates pressure-based diagnosis of instrument performance. Second, placing a frit this close to the outlet check valve causes particulates to accumulate in a location where they are likely to cause problems. Check valve failure is often caused by contaminants adhering to the ball surface, which can cause it to stick or seal improperly. In-line filtration of seal debris and other particulates downstream from the pressure transducer will minimize the chance of check valve contamination and will not complicate system diagnostics.


Pre-Column Filtration
: Two additional potential sources of particulate contamination can cause issues in the region beyond the injection device and before the analytical column. First, the injection device can introduce particles from rotor seal wear or from coring vial septa. Second, contamination may be introduced by the sample matrix. The analytical column can be damaged by irreversibly retained contaminants in the sample matrix or by salts that precipitate due to incompatibilities between the sample matrix and mobile phase. Where practical, most of these types of contamination can be addressed in the sample preparation procedure. In cases where additional filtration is required between the injector and column, a low-volume, low-dispersion filter and/or guard column must be employed. OPTI-SOLV Mini Filters and OPTI-GUARD® Guard Columns work extremely well for applications requiring pre-column protection. Our patented, automatically adjusting tube-stem design ensures zero dead volume connections and low dispersion in an area where excessive dead volume can make or break your assay.

Dead Volume and You

What Is All This Stuff About Dead Volume, Swept Volume, and Void Volumes?

This is definitely an issue of semantics. Often, two chromatographers will discuss the same concept but use different terms. Even more confusing, the word one uses will also be in the other's lexicon, but it will have a completely different meaning. To try and clear the waters, let us define some concepts and how you may hear them referred to.

Dead Volume: A difficult term to define, dead volume often refers to the volume in the flow path. However, if this definition is used, then there is no way to define the spaces in the flow path that give rise to non-linear flows. Therefore, we prefer to use the term swept volume to refer to the volume in the flow path, while ascribing the term dead volume to areas outside of the flow path that can contribute to bleeding, carryover, and cross-contamination.  

What is meant by outside the flow path? Often, gaps are introduced into the LC system, causing laminar flow to cease. These most commonly occur in connecting tubing: if tubing is not bottomed out in the tube-stop, a small space, or dead volume, is created. These small spaces serve as hiding places for samples and provide a great place for remixing analytes. This is why all Optimize products are designed to minimize dead volume. The floating stem design automatically adjusts to virtually any tube stop, making a flush zero-dead-volume (ZDV) connection, and the OPTI-LYNX™ quick-connect hardware system requires the tubing connection to be made only once, with all subsequent OPTI-LYNX connections identical.


This is not to say that minimizing swept volume is not important. Anywhere swept volume occurs is a place where mixing can happen. So, if you want to minimize mixing — which you probably do — you should also minimize the excess contributions to the swept volume. The most obvious and easiest way to do this is to use the minimum amount of tubing for connections. This is especially critical in the region between the column and the detector, where excessive volume can degrade the detector signal.

There are a few more volumes we should consider. There is the total system volume, which we have defined as swept volumeand the column volume. For column volume, there are a few definitions to examine. First, there is the amount of space in the column not taken up by packing material, which is referred to as void volume or interstitial volume. We really don't think that either is more correct, though we prefer to use interstitial volume because it is more definitive. The interstitial volume is commonly expressed temporally by the use of terms like void timecolumn dead-timet0, or tm. All of these are measurements of the time before an unretained component is detected, usually uracil. While these are the common ways to express the void volume or interstitial volume, it should be noted that these measure the contribution not only from the column, but also any extra volume from the injector through to the detector. Furthermore, they are expressed in units of time, not volume. To get volume, you must perform the favorite scientific pastime of dimensional analysis (this may excite some more than others, probably physicists).

Lastly, we should say a few words about void, or column void. This is a great example of the overuse of the word "void," which leads to a lot of confusion. When void is used by itself, it refers to a void, or space, in the actual column where there is no packing material. The presence of a void in the column is absolutely debilitating, and often spells the end of its useful life (at least as a column; it still has a very utilitarian life as a paperweight).

We hope this cleared up any confusion, both about our usage and the usage at large. It should be noted, though, that these are our definitions and you or others could use the terms above to refer to completely different things. The best way to avoid confusion in any LC conversation is to ensure that these common terms have the same meaning for all involved.

A Guideline to Effective Preventative Maintenance

As with most analytical instruments, HPLC systems require some degree of preventive maintenance to perform optimally. By proactively replacing key components before they reach their expected lifetimes, mission-critical systems can minimize unexpected and costly instrument downtime.

Service Contracts

They are certainly a convenient, if overpriced, means of handling instrument upkeep, but most routine preventative maintenance tasks can be handled quickly and easily without a visit from a service representative, even if you don’t have the benefit of an internal metrology department. By selecting easily replaced maintenance components and following proper installation procedures, you can implement your own effective preventative maintenance program. Outlined below are recommended maintenance schedules for commonly replaced components of HPLC systems. These schedules can serve as general guidelines for implementing a preventive maintenance plan for your HPLC system. Still, they can be adjusted based on your experience with maintenance cycles and typical component lifetimes in your laboratory to more closely match your operational conditions.

Piston Seals

No other wetted component within a typical HPLC system experiences as much direct wear as the piston seal. The average lifetime of a piston seal varies with several parameters, including the seal material, the composition of the mobile phase, the presence of particulates or precipitated salts around the seal, and the condition of the piston itself. Most laboratory preventive maintenance (PM) schedules call for replacement of the piston seal on a quarterly or semiannual basis, or every 3 to 6 months. In general, the piston seal should be replaced at least once every six months, provided there is no excessive and unusual wear. Where conditions are less than optimal, more frequent replacement will likely be required, such as in highly corrosive solvents or extreme operating conditions.

You can maximize seal lifetime and performance by following a few simple guidelines. First, try to pick the best seal for the mobile phase you will be using (view our seal-solvent compatibility chart for more information, or contact us for a specific recommendation for your application). Second, ensure that your piston is in good condition, is free of adhered particulates and salts, and shows no sign of wear. Finally, minimize the risk of seal damage during installation by presoaking the seal in methanol or isopropanol and using a seal-forming tool when available.

Check Valves

As with piston seals, the lifetime of a check valve will vary depending upon chromatographic conditions. The ball and seat are subject to wear, and the ball may also become coated with particulates or other matter, causing sealing problems. The majority of laboratories routinely replace check valves every six months. Once again, longer or shorter lifetimes may be the norm in your laboratory, and the frequency of replacement may require some adjustment. However, with an Optimize Technologies OPTI-MAX® Check Valve, you should get a year or more of reliable use.


When selecting a replacement check valve for your HPLC or UHPLC, you will usually have two options. For older pump models, you can use complete assemblies (or rebuild existing assemblies with multi-component rebuild kits), and for newer models, you can use a cartridge-based check valve system. Replacing whole assemblies means throwing out the entire check valve housing and internal components, then installing a new assembly. This is convenient, but possibly expensive and certainly wasteful. Rebuild kits are more economical, but the procedure is time-consuming and troublesome, and there is a risk of improper assembly and contamination. Also, some rebuild kits and assemblies incorporate a filter element within the outlet check valve. While in-line filtration is a worthwhile addition to any HPLC, putting a filter within the outlet check valve assembly has two major drawbacks. First, a filter above the outlet ball will concentrate particulates in a region where they can potentially cause check valve failure. Second, the presence of an in-line filter element upstream from the pressure transducer can complicate pressure-based instrument diagnostics.

Cartridge-based systems are the preferred choice due to greater convenience and reliability, greater economy, reduced waste, and a lower risk of contamination during installation. When selecting cartridge check valve products, focus on manufacturing quality and product performance, and beware of dubious marketing claims. For instance, several manufacturers claim that their check valve cartridges are “self-priming.” This carries little importance or credence, as check valves are, by design, passive flow-restriction devices – they are self-priming in the same way that rivers are self-flowing. Positive hydrostatic pressure causes the mobile phase to flow through ball-and-seat-type check valves – as long as there is minimal resistance to flow in the forward direction, a check valve won’t get in the way of the mobile phase. The specifications that really matter for check valve performance include ball/seat material, finish quality, leak rating, and the allowable ball travel.

Pistons

The pistons inside an HPLC pump do not require routine replacement, but they too are subject to wear and should be inspected as part of the PM schedule. Pistons should be examined at least once a year and whenever the seal is replaced. Close examination of the surface finish can reveal potential problems such as flat spots, out-of-roundness, adhered particulates or buffer salts, scratches, and cracks. Light microscopes can be useful for inspecting the condition of a piston.

Particular attention should be paid to the piston when UHMW-PE seals are to be installed in place of PTFE-based seals. UHMW-PE provides considerably longer seal lifetimes when used with mostly aqueous mobile phases, but the PE material is less forgiving than the softer PTFE blend. A worn or damaged piston that performs adequately with a PTFE seal may cause problems with UHMW-PE. 

When switching from PTFE to UHMW-PE seals, it is advisable to replace the piston at the same time.

UV Detectors

For UV detectors, source lamps are the most common consumable item requiring routine replacement. The replacement frequency will vary with usage, but it can be estimated based on usage and the lamp's anticipated lifetime. Typically, source lamps will require replacement annually or semi-annually. Some types of lamps come pre-aligned in holders, ready for immediate installation. If pre-aligned lamps are not available for your detector, alignment will be required during installation. This can be a somewhat involved procedure, and you may wish to consult with qualified personnel if you have not successfully performed an alignment before.

Autosampler or Injection Valve

For most sample injection devices, the only components that require regular replacement are rotor seals and liquid seals. This can usually be performed annually.

Conclusion

In a productive lab, the cost of HPLC downtime can be thousands of dollars for each hour the system is down. Implementing an effective preventative maintenance program for your HPLC can help prevent minor maintenance issues from becoming a major drain on your time and budget.

The Facts about Replacement HPLC Components

Weighing Options with the Facts in Hand 

Before you purchased your last HPLC system, you probably spent plenty of time researching your options. You pored over ads and brochures, consulted with colleagues, and met with the sales and support personnel from prospective instrument manufacturers - all to make sure that you would be buying an instrument well-suited for your needs, from a company that makes a point of standing behind both the equipment it makes and the customers who buy it. Now that this HPLC system has been in service for a year or two, you are likely weighing your options for suppliers of replacement components. The selection criteria should be similar to those used in purchasing the HPLC system itself. After all, you depend on these components to function reliably and smoothly in your HPLC, so you can concentrate on your analysis without costly, unexpected interruptions in throughput. The list of potential parts suppliers is shorter than that for suppliers of new HPLC instrumentation, but the choice requires the same level of scrutiny. The decision will have a significant impact on operational costs and the time you spend replacing vital components and diagnosing system problems.

OEM Components: Fact and Fiction

With a new HPLC instrument, your first set of replacement parts will be supplied by the original equipment manufacturer (OEM). In fact, due to instrument warranty restrictions, it’s likely that for a period of time you’ll be somewhat of a captive market for the OEM when it comes to purchasing components. OEM instrument companies want customers to buy only components that meet their performance specifications, on the grounds that the original manufacturer knows best which components will function optimally. But they also want customers to buy only the components they sell. The reality is that few OEMs actually make replacement components themselves; much of this work is farmed out to other manufacturers. The quality and reliability of OEM components will vary; the products are fairly reliable, but command a premium price. Also, OEMs are rarely a driving force behind product innovation at the replacement-component level, and, in fact, some of our most popular products were developed specifically to address shortcomings in original equipment!

Third Party Suppliers

Third-party suppliers, like Optimize Technologies, are a second option for sourcing components and accessories. Unlike Optimize, most of these companies offer only their own versions of standard OEM components, such as full check valve assemblies and rebuild kits, and differentiate themselves solely on price. There are rarely any distinct advantages in either product design or performance. 

The Optimize Difference 

At Optimize Technologies, we continually strive to be the exception among third-party parts suppliers. Since 1985, Optimize has been manufacturing HPLC components and accessories with two goals: to produce the most innovative and reliable components available and to stand behind them with the most responsive and effective customer service in the industry.

Every component we manufacture is designed with an important purpose in mind: to keep your HPLC system operating at peak performance between scheduled maintenance periods. We precision-machine our OEM replacement components to the most exacting standards in the industry. Balls and seats used in our OPTI-MAX® cartridge check valves are precisely matched to meet leak-test benchmarks four times more stringent than the industry standard. OPTI-SEAL® and ITB™ piston seals are engineered to provide longer lifetimes and superior shedding characteristics. Replacement pistons from Optimize offer excellent concentricity and consistent surface finish. Whether you are buying a fitting, a priming valve, or a check valve conversion kit from Optimize, you can always be sure that our commitment to quality and innovation is paramount.

Most instrument manufacturers will tell you quite emphatically that for optimal instrument performance, you should use only components that meet their performance specifications. At Optimize, we couldn't agree more. Every replacement component we offer is fully guaranteed to meet or exceed the specifications of the original equipment manufacturer. In fact, Optimize Technologies already produces HPLC components for several OEM manufacturers, like Agilent and Waters, which are incorporated as original equipment on new HPLC instrumentation.

Value by Design

What parameters really affect the performance of a fluid-handling component in your HPLC system? Which seal polymer blends deliver optimal wear resistance and longer lifetimes? What factors are important in producing a check valve that performs optimally in any instrument? At Optimize, research into these types of questions is an ongoing activity. Our investment in R&D provides us with the information we need to optimize manufacturing specifications, and our laboratory experience gives us unique insight into the challenges you face in instrument metrology. The result is a line of HPLC components well-suited to deliver the performance you demand, designed with innovative, time-saving features that streamline common maintenance tasks and ensure trouble-free installation and instrument qualification.

We understand that one of the costliest resources in a productive laboratory is time. When an HPLC system is rendered inoperative due to a maintenance problem, the opportunity cost of lost analysis time can add up very quickly, potentially far exceeding the cost of the parts that caused the problem to begin with. By offering only components of superior quality and unsurpassed reliability, we can help you spend less time on instrument maintenance and more time on what’s really important – your chromatography.

Your Resource for New Innovation

At Optimize Technologies, we offer a comprehensive selection of components for most HPLC brands, and we're always seeking new ways to deliver products that save your laboratory time and money. If you need a product not listed in our catalog or online, please call a technical support representative to discuss your specific fluid-handling or overall HPLC analysis challenges.